Practical Radiation Dose Using Modern and Aged Angiographic Equipment

Main Article Content

Taysia P. T. Morioka, BSE
Alisha E. Suri, BSc
Jennifer Snyder, MM
Casey Yara, AS
Scott B. Raymond, MD, PhD
Yi Zhang, MD
Samuel Tsappidi, MD
Ferdinand K. Hui, MD

Abstract

Background: Replacement of angiographic capital equipment is costly, particularly in resource-limited settings where aging platforms remain in clinical use. Advances in detector technology, image processing, and automated dose control have substantially reduced radiation exposure in modern systems, yet real-world generational comparisons remain limited.


Methods: This retrospective single-center study evaluated 363 neurointerventional procedures performed between September 2023 and February 2024 at The Queen’s Medical Center. Cases were completed on either a 17-year-old Philips Allura Xper FD10/10 (n=189) or a 1-year-old Siemens Artis Icono (n=174). Procedures were categorized as diagnostic angiography, embolization, or stroke intervention. Radiation exposure was assessed using system-reported cumulative air kerma. Medians and interquartile ranges were compared using Wilcoxon rank-sum testing.


Results: Across all procedures, median cumulative air kerma was significantly higher on the older system (1267 [902–1827] mGy) compared with the newer platform (410 [251–677] mGy; p<0.001), representing an approximately three-fold reduction following modernization. Diagnostic angiography demonstrated the largest relative decrease (1085 vs 280 mGy), followed by embolization (1847 vs 639 mGy) and stroke intervention (1382 vs 371 mGy) (all p<0.001). Clinically relevant thresholds were exceeded more frequently on the older system: 17.5% of cases surpassed 2 Gy versus 0.6% on the newer platform, while no procedures on the modern system exceeded 3 Gy.


Conclusions: Contemporary angiographic technology was associated with three- to four-fold reductions in radiation exposure across neurointerventional procedures, underscoring the substantial safety benefits of equipment modernization.


Keywords: Radiation dose, biplane angiography, air kerma, equipment modernization, patient safety.


Abbreviations: ALARA, as low as reasonably achievable; Gy, gray; mGy, milligray; IQR, interquartile range.

Downloads

Download data is not yet available.

Article Details

How to Cite
Morioka, T. P. T., Suri, A. E., Snyder, J., Yara, C., Raymond, S. B., Zhang, Y., Tsappidi, S., & Hui, F. K. (2026). Practical Radiation Dose Using Modern and Aged Angiographic Equipment. Journal of Vascular and Interventional Neurology, 16(3), 112–117. Retrieved from https://jvin.org/index.php/jvin/article/view/504